*** MIDAS logfile (03SEPpl1.2) *** Mo, 28 Feb 2005 17:44:10 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 28 Feb 2005 17:45:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050228 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 352.48 stddev = 5.57 FIERA temperature = 291.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050228 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1072.1 1071.9 895.5 -0.01 -0.01 1.20 2 2.00 2057.4 2058.0 1749.7 -0.01 -0.00 1.18 3 4.00 4027.8 4027.3 3233.6 0.01 0.00 1.25 4 6.00 5998.5 5998.4 4792.4 0.02 0.02 1.25 5 9.00 8952.6 8953.6 7111.8 0.01 0.02 1.26 6 12.00 11908.2 11907.9 9506.3 0.02 0.01 1.25 7 15.00 14862.1 14863.5 11836.4 0.02 0.02 1.26 8 20.00 19786.8 19786.4 15905.3 0.01 0.01 1.24 9 25.00 24703.3 24703.8 19525.7 -0.02 -0.02 1.27 10 30.00 29616.3 29616.4 23028.2 -0.05 -0.05 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1370 to 37861 e- Average count rate measured: 985 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.01 e-/ADU RON = 7.12 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.091 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay