*** MIDAS logfile (03SEPpl1.2) *** Su, 20 Feb 2005 22:34:43 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 20 Feb 2005 22:35:55 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 353.90 stddev = 5.84 FIERA temperature = 291.4 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1018.1 1017.3 853.7 0.03 0.00 1.19 2 2.00 1964.9 1965.3 1634.0 -0.03 0.01 1.20 3 4.00 3860.0 3859.9 3098.6 -0.00 -0.01 1.25 4 6.00 5755.0 5755.8 4640.1 0.00 0.01 1.24 5 9.00 8597.6 8597.2 6883.8 0.01 0.01 1.25 6 12.00 11439.5 11439.9 9225.7 0.01 0.02 1.24 7 15.00 14282.2 14281.9 11416.1 0.02 0.02 1.25 8 20.00 19017.3 19016.4 14993.0 0.01 0.00 1.27 9 25.00 23746.4 23748.2 18636.0 -0.02 -0.01 1.27 10 30.00 28473.7 28474.7 22319.9 -0.05 -0.05 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1303 to 36443 e- Average count rate measured: 947 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.00 e-/ADU RON = 7.48 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.076 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay