*** MIDAS logfile (03SEPpl1.2) *** Sa, 05 Feb 2005 15:59:28 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 05 Feb 2005 16:00:33 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050205 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 352.49 stddev = 5.43 FIERA temperature = 291.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050205 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1075.0 1075.5 1089.1 -0.03 -0.02 0.99 2 2.00 2062.9 2062.6 1687.0 -0.01 -0.02 1.22 3 4.00 4037.7 4039.1 4450.2 0.00 0.02 0.91 4 6.00 6012.7 6012.8 4851.0 0.01 0.01 1.24 5 9.00 8975.2 8975.2 7209.1 0.02 0.01 1.24 6 12.00 11937.8 11937.7 9544.0 0.02 0.02 1.25 7 15.00 14900.1 14900.1 11950.2 0.02 0.02 1.25 8 20.00 19834.4 19836.3 15781.0 0.01 0.02 1.26 9 25.00 24763.6 24765.8 19564.2 -0.01 -0.01 1.27 10 30.00 29688.8 29689.5 23190.5 -0.05 -0.05 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1398 to 38596 e- Average count rate measured: 987 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.02 e-/ADU RON = 7.06 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.092 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay