*** MIDAS logfile (03SEPpl1.2) *** Su, 23 Jan 2005 18:24:26 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 23 Jan 2005 18:25:33 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050123 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.01 stddev = 7.54 FIERA temperature = 291.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050123 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1008.3 1007.3 870.9 -0.02 -0.00 1.16 2 2.00 1948.5 1948.6 1614.6 -0.04 -0.01 1.21 3 4.00 3829.9 3830.2 3098.2 -0.01 0.00 1.24 4 6.00 5712.0 5711.5 4602.7 0.01 0.00 1.24 5 9.00 8534.7 8534.6 6870.9 0.03 0.02 1.24 6 12.00 11356.8 11356.2 9120.9 0.03 0.01 1.25 7 15.00 14179.3 14179.9 11359.1 0.03 0.03 1.25 8 20.00 18880.2 18881.0 14972.0 0.02 0.01 1.26 9 25.00 23579.4 23578.9 18724.4 0.00 -0.01 1.26 10 30.00 28269.6 28270.9 22394.5 -0.04 -0.05 1.26 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1276 to 35795 e- Average count rate measured: 940 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.00 e-/ADU RON = 9.54 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.073 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay