*** MIDAS logfile (03SEPpl1.2) *** Sa, 15 Jan 2005 16:51:35 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 15 Jan 2005 16:52:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050115 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.06 stddev = 7.13 FIERA temperature = 290.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050115 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 994.2 996.8 851.5 -0.03 0.03 1.17 2 2.00 1938.2 1937.8 1626.8 -0.02 -0.01 1.19 3 4.00 3820.8 3821.3 3122.6 0.00 0.01 1.22 4 6.00 5704.0 5703.6 4581.4 0.03 0.01 1.25 5 9.00 8525.8 8526.4 6774.4 0.01 0.01 1.26 6 12.00 11350.1 11351.0 9092.0 0.03 0.02 1.25 7 15.00 14173.1 14173.4 11242.7 0.03 0.01 1.26 8 20.00 18875.4 18877.4 15201.0 0.01 0.01 1.24 9 25.00 23574.8 23575.4 19211.6 -0.01 -0.02 1.23 10 30.00 28268.8 28270.2 22216.3 -0.04 -0.06 1.27 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1252 to 35549 e- Average count rate measured: 941 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.26 +/- 0.01 e-/ADU RON = 8.96 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.061 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay