*** MIDAS logfile (03SEPpl1.2) *** Fr, 07 Jan 2005 14:51:01 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 07 Jan 2005 14:52:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050107 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.20 stddev = 5.39 FIERA temperature = 290.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050107 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1022.4 1022.1 876.0 0.01 -0.00 1.17 2 2.00 1974.9 1974.8 1691.5 -0.01 -0.01 1.17 3 4.00 3880.9 3880.1 3102.8 0.02 0.00 1.25 4 6.00 5785.8 5785.6 4630.3 0.01 0.01 1.25 5 9.00 8643.3 8643.1 6916.4 0.01 0.02 1.25 6 12.00 11501.2 11500.7 9152.2 0.02 0.02 1.26 7 15.00 14358.4 14357.7 11419.0 0.02 0.02 1.26 8 20.00 19118.0 19119.4 15142.5 0.00 0.01 1.26 9 25.00 23874.8 23874.9 18897.1 -0.02 -0.02 1.26 10 30.00 28626.5 28626.6 22441.8 -0.05 -0.05 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1304 to 36522 e- Average count rate measured: 952 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.00 e-/ADU RON = 6.87 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.075 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay