*** MIDAS logfile (03SEPpl1.2) *** Sa, 01 Jan 2005 17:27:17 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 01 Jan 2005 17:28:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050101 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.80 stddev = 5.87 FIERA temperature = 291.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050101 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1085.8 1086.0 1100.1 0.02 0.03 0.99 2 2.00 2085.0 2084.8 1709.7 0.01 0.01 1.22 3 4.00 4080.6 4080.8 3287.4 -0.00 0.00 1.24 4 6.00 6079.0 6078.4 4952.3 0.02 0.01 1.23 5 9.00 9073.9 9073.0 8096.0 0.02 0.01 1.12 6 12.00 12068.9 12069.1 9678.6 0.02 0.02 1.25 7 15.00 15063.6 15062.5 12405.5 0.02 0.01 1.21 8 20.00 20051.6 20052.5 15938.6 -0.00 0.00 1.26 9 25.00 25033.8 25034.6 19494.2 -0.03 -0.03 1.28 10 30.00 30014.5 30015.3 23344.5 -0.06 -0.06 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1407 to 38881 e- Average count rate measured: 998 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.02 e-/ADU RON = 7.60 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.090 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay