*** MIDAS logfile (05SEPpl1.2) *** We, 19 Sep 2007 17:53:02 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 19 Sep 2007 17:53:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 070919 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 257.33 stddev = 4.33 FIERA temperature = 282.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_070919 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1707.8 1707.9 1377.5 -0.08 -0.10 1.24 2 8.51 3378.1 3378.0 2702.8 0.07 0.07 1.25 3 17.02 6714.1 6714.4 5321.2 0.10 0.10 1.26 4 25.53 10050.0 10050.6 7840.4 0.11 0.12 1.28 5 38.29 15055.5 15055.2 11535.2 0.13 0.13 1.31 6 51.06 20052.2 20052.4 15203.6 0.09 0.10 1.32 7 63.83 25048.3 25050.1 18645.7 0.07 0.08 1.34 8 85.10 33357.8 33357.6 24064.0 -0.01 -0.00 1.39 9 106.38 41625.1 41624.2 29213.9 -0.15 -0.15 1.42 10 127.66 49849.1 49850.1 34012.3 -0.34 -0.33 1.47 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2502 to 73037 e- Average count rate measured: 391 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.47 +/- 0.03 e-/ADU RON = 6.34 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.116 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.34 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay