*** MIDAS logfile (05SEPpl1.2) *** We, 05 Sep 2007 16:29:33 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 05 Sep 2007 16:29:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 070905 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 266.52 stddev = 4.35 FIERA temperature = 282.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_070905 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1637.7 1637.7 1330.5 -0.08 -0.09 1.23 2 8.51 3251.0 3250.8 2607.9 0.07 0.06 1.25 3 17.02 6473.3 6473.4 5142.9 0.09 0.10 1.26 4 25.53 9696.1 9695.9 7591.0 0.11 0.11 1.28 5 38.29 14530.5 14530.2 11195.5 0.11 0.12 1.30 6 51.06 19358.4 19358.7 14734.0 0.09 0.09 1.31 7 63.83 24186.1 24185.5 18138.7 0.07 0.07 1.33 8 85.10 32213.7 32214.4 23370.9 -0.01 -0.00 1.38 9 106.38 40204.5 40205.9 28744.3 -0.14 -0.14 1.40 10 127.66 48160.8 48160.9 33143.0 -0.31 -0.31 1.45 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2370 to 69686 e- Average count rate measured: 378 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.45 +/- 0.02 e-/ADU RON = 6.29 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.083 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.31 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay