*** MIDAS logfile (05SEPpl1.2) *** Su, 01 Jul 2007 19:03:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 01 Jul 2007 19:03:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 070701 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 265.52 stddev = 5.23 FIERA temperature = 284.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_070701 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1745.4 1746.0 1444.2 -0.10 -0.11 1.21 2 8.51 3455.3 3455.0 2777.2 0.09 0.07 1.24 3 17.02 6869.1 6869.1 5438.2 0.12 0.12 1.26 4 25.53 10282.8 10282.7 8069.9 0.12 0.13 1.27 5 38.30 15404.1 15404.0 11890.9 0.13 0.14 1.30 6 51.06 20517.3 20517.8 15640.8 0.10 0.10 1.31 7 63.83 25629.8 25630.4 19056.9 0.08 0.08 1.34 8 85.10 34126.9 34127.2 24695.9 -0.02 -0.01 1.38 9 106.38 42584.6 42583.4 29735.8 -0.17 -0.16 1.43 10 127.66 50992.9 50990.9 34441.3 -0.36 -0.36 1.48 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2579 to 75339 e- Average count rate measured: 401 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.48 +/- 0.03 e-/ADU RON = 7.73 +/- 0.15 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.110 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.36 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay