*** MIDAS logfile (05SEPpl1.2) *** Sa, 05 Aug 2006 15:59:56 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 05 Aug 2006 16:01:10 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 060805 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 227.04 stddev = 5.14 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_060805 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1812.1 1812.3 1453.3 -0.06 -0.06 1.25 2 8.51 3586.3 3586.4 2836.6 0.05 0.05 1.26 3 17.02 7131.5 7131.7 5645.6 0.07 0.07 1.26 4 25.53 10677.6 10677.8 8196.2 0.09 0.09 1.30 5 38.30 15999.0 15999.3 12255.7 0.11 0.11 1.31 6 51.06 21314.6 21314.3 15782.2 0.10 0.09 1.35 7 63.83 26630.5 26630.9 19470.7 0.09 0.09 1.37 8 85.10 35465.1 35466.9 25194.7 0.01 0.01 1.41 9 106.38 44257.0 44258.2 30218.2 -0.13 -0.13 1.46 10 127.66 53005.3 53003.6 35133.6 -0.31 -0.32 1.51 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2734 to 79966 e- Average count rate measured: 416 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.51 +/- 0.03 e-/ADU RON = 7.75 +/- 0.15 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.105 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.31 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay