*** MIDAS logfile (05SEPpl1.2) *** Th, 06 Apr 2006 14:43:04 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 06 Apr 2006 14:44:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 060406 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 267.84 stddev = 4.96 FIERA temperature = 289.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_060406 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1876.7 1876.7 1517.6 -0.11 -0.12 1.24 2 8.51 3718.2 3718.0 2960.1 0.08 0.08 1.26 3 17.02 7396.2 7395.8 5816.2 0.14 0.13 1.27 4 25.53 11073.5 11074.2 8618.7 0.14 0.15 1.28 5 38.30 16590.0 16589.6 12896.3 0.15 0.15 1.29 6 51.06 22098.6 22099.1 16540.7 0.12 0.12 1.34 7 63.83 27605.0 27604.6 20275.3 0.09 0.09 1.36 8 85.10 36752.2 36750.9 26367.0 -0.01 -0.02 1.39 9 106.38 45848.4 45847.6 31431.2 -0.19 -0.19 1.46 10 127.66 54885.9 54886.3 36644.7 -0.41 -0.41 1.50 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2813 to 82275 e- Average count rate measured: 431 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.50 +/- 0.03 e-/ADU RON = 7.43 +/- 0.15 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.103 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.41 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay