*** MIDAS logfile (05SEPpl1.2) *** We, 29 Mar 2006 15:42:19 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 29 Mar 2006 15:43:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 060329 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 256.10 stddev = 5.37 FIERA temperature = 289.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_060329 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1867.7 1867.6 1518.5 -0.11 -0.11 1.23 2 8.51 3700.1 3699.8 2944.7 0.09 0.08 1.26 3 17.02 7359.7 7359.5 5821.2 0.14 0.14 1.26 4 25.53 11017.8 11017.8 8660.2 0.14 0.14 1.27 5 38.30 16507.3 16506.1 12653.4 0.15 0.15 1.30 6 51.06 21987.9 21986.4 16465.6 0.12 0.11 1.34 7 63.83 27467.5 27467.5 20348.7 0.09 0.10 1.35 8 85.10 36565.6 36565.7 26023.9 -0.02 -0.02 1.41 9 106.38 45625.4 45623.5 31494.3 -0.18 -0.18 1.45 10 127.66 54610.4 54609.8 36403.6 -0.42 -0.42 1.50 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2799 to 81840 e- Average count rate measured: 429 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.50 +/- 0.03 e-/ADU RON = 8.05 +/- 0.16 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.104 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.42 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay