*** MIDAS logfile (05SEPpl1.2) *** Th, 16 Feb 2006 17:20:28 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 16 Feb 2006 17:21:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 060216 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 264.44 stddev = 4.52 FIERA temperature = 289.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_060216 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1915.6 1915.1 1560.7 -0.12 -0.13 1.23 2 8.51 3792.6 3792.7 3037.6 0.08 0.09 1.25 3 17.02 7543.7 7543.3 5946.1 0.14 0.14 1.27 4 25.53 11293.2 11293.3 8763.9 0.15 0.15 1.29 5 38.30 16918.1 16917.9 12921.6 0.16 0.16 1.31 6 51.06 22535.2 22534.9 16792.7 0.13 0.13 1.34 7 63.83 28148.8 28147.9 20618.0 0.10 0.10 1.37 8 85.10 37471.8 37471.1 26725.3 -0.02 -0.02 1.40 9 106.38 46741.4 46741.9 32211.9 -0.20 -0.20 1.45 10 127.66 55950.8 55952.1 37050.3 -0.43 -0.43 1.51 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2882 to 84180 e- Average count rate measured: 440 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.50 +/- 0.03 e-/ADU RON = 6.80 +/- 0.14 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.107 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.43 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay