*** MIDAS logfile (05SEPpl1.2) *** Mo, 13 Feb 2006 18:45:38 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 13 Feb 2006 18:46:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 060213 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 264.44 stddev = 5.16 FIERA temperature = 289.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_060213 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 955.9 956.4 794.2 -0.00 0.01 1.20 2 8.00 1890.6 1890.6 1551.3 -0.02 -0.02 1.22 3 16.00 3760.7 3760.7 3003.8 0.01 0.01 1.25 4 24.00 5630.6 5630.9 4558.7 0.02 0.02 1.24 5 36.00 8434.9 8435.0 6614.9 0.01 0.01 1.28 6 48.00 11240.3 11240.4 8771.7 0.03 0.02 1.28 7 60.00 14044.2 14045.1 10851.2 0.02 0.02 1.29 8 80.00 18715.2 18716.2 14166.3 0.00 0.01 1.32 9 100.00 23382.8 23383.0 17496.1 -0.02 -0.02 1.34 10 120.00 28046.9 28045.7 20676.0 -0.05 -0.06 1.36 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1300 to 38130 e- Average count rate measured: 234 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.36 +/- 0.01 e-/ADU RON = 7.01 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.095 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay