*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 17:44:19 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 17:45:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 060114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 273.84 stddev = 3.73 FIERA temperature = 291.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 922.7 922.5 763.7 0.00 -0.01 1.21 2 8.00 1828.1 1827.9 1471.4 -0.01 -0.01 1.24 3 16.00 3638.9 3639.0 2924.9 0.01 0.01 1.24 4 24.00 5449.4 5449.5 4322.1 0.01 0.01 1.26 5 36.00 8166.1 8166.1 6417.0 0.02 0.02 1.27 6 48.00 10882.8 10882.8 8449.4 0.02 0.02 1.29 7 60.00 13598.3 13599.0 10703.0 0.02 0.02 1.27 8 80.00 18123.2 18123.5 13798.5 0.00 0.00 1.31 9 100.00 22644.4 22644.3 16986.4 -0.02 -0.02 1.33 10 120.00 27161.9 27162.0 20093.4 -0.05 -0.05 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1247 to 36705 e- Average count rate measured: 226 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.01 e-/ADU RON = 5.04 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.079 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay