*** MIDAS logfile (05SEPpl1.2) *** Mo, 26 Dec 2005 19:26:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 26 Dec 2005 19:28:10 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 051226 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 265.90 stddev = 5.42 FIERA temperature = 291.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_051226 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 913.3 913.0 760.8 -0.00 -0.00 1.20 2 8.00 1812.0 1811.8 1481.3 -0.01 -0.01 1.22 3 16.00 3610.5 3610.1 2902.7 0.01 0.01 1.24 4 24.00 5408.4 5408.2 4285.1 0.01 0.01 1.26 5 36.00 8105.6 8105.6 6541.8 0.02 0.02 1.24 6 48.00 10802.7 10802.9 8448.7 0.02 0.02 1.28 7 60.00 13499.6 13499.6 10571.1 0.02 0.02 1.28 8 80.00 17991.8 17992.3 13750.0 0.00 0.00 1.31 9 100.00 22481.8 22482.2 16854.9 -0.02 -0.02 1.33 10 120.00 26966.8 26966.3 20036.3 -0.05 -0.05 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1232 to 36393 e- Average count rate measured: 225 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.01 e-/ADU RON = 7.32 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.066 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay