*** MIDAS logfile (03SEPpl1.2) *** Th, 20 Oct 2005 14:46:06 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Th, 20 Oct 2005 14:47:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 051020 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 281.09 stddev = 4.03 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_051020 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 978.1 977.3 799.2 -0.00 -0.01 1.22 2 8.00 1933.9 1934.2 1563.8 -0.01 0.02 1.24 3 16.00 3846.9 3846.7 3112.5 0.00 0.00 1.24 4 24.00 5760.5 5760.5 4600.5 0.02 0.02 1.25 5 36.00 8630.7 8630.6 6775.6 0.02 0.02 1.27 6 48.00 11501.4 11500.8 9062.0 0.03 0.02 1.27 7 60.00 14369.5 14369.9 11096.9 0.02 0.02 1.29 8 80.00 19149.7 19150.6 14648.6 0.01 0.00 1.31 9 100.00 23924.7 23925.5 17841.4 -0.02 -0.03 1.34 10 120.00 28697.3 28697.3 21070.7 -0.05 -0.06 1.36 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1331 to 39059 e- Average count rate measured: 239 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.36 +/- 0.01 e-/ADU RON = 5.48 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.092 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay