*** MIDAS logfile (03SEPpl1.2) *** Sa, 08 Oct 2005 12:43:29 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 08 Oct 2005 12:44:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 051008 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 280.80 stddev = 4.40 FIERA temperature = 290.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_051008 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 975.8 975.6 800.5 -0.01 -0.01 1.22 2 8.00 1930.5 1930.5 1565.9 0.00 0.01 1.23 3 16.00 3839.4 3839.2 3074.2 -0.00 -0.00 1.25 4 24.00 5749.3 5749.5 4551.1 0.01 0.02 1.26 5 36.00 8613.9 8613.7 6951.1 0.02 0.02 1.24 6 48.00 11478.6 11478.5 8952.1 0.03 0.02 1.28 7 60.00 14341.7 14342.2 11079.9 0.02 0.02 1.29 8 80.00 19113.0 19113.9 14531.3 0.01 0.01 1.32 9 100.00 23879.2 23879.2 17882.0 -0.02 -0.02 1.34 10 120.00 28641.4 28642.6 21061.1 -0.05 -0.05 1.36 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1327 to 38943 e- Average count rate measured: 239 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.36 +/- 0.01 e-/ADU RON = 5.98 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.092 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay