*** MIDAS logfile (03SEPpl1.2) *** We, 14 Sep 2005 16:29:00 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 14 Sep 2005 16:30:19 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 050914 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 279.59 stddev = 4.08 FIERA temperature = 289.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_050914 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 940.8 940.9 770.1 -0.02 -0.01 1.22 2 8.00 1864.0 1863.7 1500.7 0.00 0.00 1.24 3 16.00 3708.7 3709.0 2954.8 -0.00 0.01 1.26 4 24.00 5554.8 5554.9 4404.3 0.02 0.02 1.26 5 36.00 8322.6 8322.5 6563.0 0.02 0.01 1.27 6 48.00 11091.6 11091.2 8613.5 0.02 0.02 1.29 7 60.00 13858.4 13859.5 10825.0 0.02 0.02 1.28 8 80.00 18470.0 18469.5 14113.0 0.01 0.00 1.31 9 100.00 23076.1 23077.4 17258.7 -0.02 -0.02 1.34 10 120.00 27680.2 27679.9 20402.7 -0.05 -0.05 1.36 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1275 to 37500 e- Average count rate measured: 231 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.01 e-/ADU RON = 5.53 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.081 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay