*** MIDAS logfile (03SEPpl1.2) *** Su, 04 Sep 2005 16:40:53 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 04 Sep 2005 16:42:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 050904 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 280.41 stddev = 5.00 FIERA temperature = 290.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_050904 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 922.6 922.7 763.9 -0.01 -0.01 1.21 2 8.00 1832.0 1831.6 1490.0 0.01 -0.01 1.23 3 16.00 3649.5 3649.5 2927.2 0.01 0.01 1.25 4 24.00 5467.3 5467.7 4337.1 0.01 0.01 1.26 5 36.00 8194.6 8194.4 6429.9 0.02 0.02 1.27 6 48.00 10921.7 10922.1 8542.6 0.02 0.02 1.28 7 60.00 13648.0 13648.4 10594.8 0.01 0.02 1.29 8 80.00 18190.7 18190.8 13841.4 0.00 0.00 1.31 9 100.00 22730.2 22731.0 17035.5 -0.02 -0.01 1.33 10 120.00 27264.8 27264.4 20265.8 -0.05 -0.05 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1245 to 36786 e- Average count rate measured: 227 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.01 e-/ADU RON = 6.75 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.062 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay