*** MIDAS logfile (03SEPpl1.2) *** We, 29 Jun 2005 20:31:15 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 29 Jun 2005 20:32:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 050629 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 282.07 stddev = 6.54 FIERA temperature = 290.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_050629 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 983.5 983.5 835.5 -0.01 -0.00 1.18 2 8.00 1944.0 1943.9 1594.4 -0.00 0.00 1.22 3 16.00 3864.6 3864.3 3104.3 0.01 0.00 1.24 4 24.00 5785.5 5785.6 4641.1 0.01 0.02 1.25 5 36.00 8666.9 8667.0 6912.0 0.02 0.02 1.25 6 48.00 11548.2 11548.1 9004.4 0.02 0.02 1.28 7 60.00 14428.3 14429.0 11125.0 0.02 0.02 1.30 8 80.00 19227.8 19227.1 14585.7 0.01 0.00 1.32 9 100.00 24022.7 24023.1 18251.0 -0.02 -0.02 1.32 10 120.00 28811.6 28811.1 21318.8 -0.06 -0.06 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1328 to 38888 e- Average count rate measured: 240 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.01 e-/ADU RON = 8.83 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.100 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay