*** MIDAS logfile (03SEPpl1.2) *** Mo, 28 Feb 2005 16:20:28 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 28 Feb 2005 16:21:55 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 050228 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.36 stddev = 5.55 FIERA temperature = 291.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_050228 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 1003.1 1003.1 835.5 0.00 -0.00 1.20 2 8.00 1984.6 1984.6 1692.9 0.00 -0.00 1.17 3 16.00 3947.1 3947.2 3139.5 -0.01 -0.00 1.26 4 24.00 5910.8 5910.6 4675.3 0.01 0.01 1.26 5 36.00 8855.9 8855.9 7047.5 0.02 0.02 1.26 6 48.00 11801.2 11801.0 9125.8 0.02 0.02 1.29 7 60.00 14745.7 14745.7 11326.9 0.02 0.02 1.30 8 80.00 19650.6 19651.7 14788.8 0.00 0.01 1.33 9 100.00 24552.5 24551.8 18216.7 -0.02 -0.02 1.35 10 120.00 29449.0 29449.1 21567.7 -0.05 -0.05 1.37 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1375 to 40380 e- Average count rate measured: 245 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.37 +/- 0.01 e-/ADU RON = 7.61 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.090 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay