*** MIDAS logfile (03SEPpl1.2) *** Sa, 01 Jan 2005 16:50:57 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 01 Jan 2005 16:52:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 050101 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 276.04 stddev = 5.38 FIERA temperature = 291.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_050101 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 1016.4 1016.5 840.2 -0.01 -0.01 1.21 2 8.00 2011.6 2011.6 1630.7 0.00 0.00 1.23 3 16.00 4001.0 4001.0 3189.3 -0.00 -0.00 1.25 4 24.00 5991.4 5991.2 4736.4 0.01 0.01 1.26 5 36.00 8977.0 8976.5 7023.2 0.02 0.02 1.28 6 48.00 11961.8 11961.9 9254.8 0.02 0.02 1.29 7 60.00 14946.7 14946.5 11457.0 0.02 0.02 1.30 8 80.00 19918.5 19919.2 15032.7 0.01 0.01 1.33 9 100.00 24887.4 24886.8 18515.9 -0.02 -0.02 1.34 10 120.00 29847.9 29848.0 21762.5 -0.06 -0.06 1.37 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1393 to 40894 e- Average count rate measured: 249 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.37 +/- 0.01 e-/ADU RON = 7.38 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.090 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay