*** MIDAS logfile (03SEPpl1.2) *** Mo, 20 Dec 2004 17:14:51 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 20 Dec 2004 17:16:16 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 041220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.84 stddev = 5.37 FIERA temperature = 290.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_041220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 1010.5 1010.2 885.3 -0.01 -0.01 1.14 2 8.00 1998.7 1998.5 1610.7 0.00 0.00 1.24 3 16.00 3974.4 3974.5 3160.5 -0.00 -0.00 1.26 4 24.00 5951.4 5951.0 4755.3 0.01 0.01 1.25 5 36.00 8915.8 8915.9 6970.5 0.01 0.02 1.28 6 48.00 11881.2 11881.2 9204.1 0.02 0.02 1.29 7 60.00 14846.0 14845.5 11537.7 0.02 0.02 1.29 8 80.00 19784.1 19783.9 14879.1 0.01 0.01 1.33 9 100.00 24718.2 24718.5 18803.7 -0.02 -0.02 1.31 10 120.00 29646.2 29645.3 21562.2 -0.05 -0.06 1.37 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1377 to 40393 e- Average count rate measured: 247 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.36 +/- 0.02 e-/ADU RON = 7.32 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.092 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay