*** MIDAS logfile (03SEPpl1.2) *** We, 06 Oct 2004 15:37:48 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 06 Oct 2004 15:39:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 041006 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.69 stddev = 5.13 FIERA temperature = 290.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_041006 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 1027.1 1027.2 844.6 -0.02 -0.01 1.22 2 8.00 2032.1 2032.0 1667.2 0.01 0.01 1.22 3 16.00 4040.3 4040.0 3259.8 0.00 -0.00 1.24 4 24.00 6049.2 6049.6 4781.9 0.01 0.01 1.27 5 36.00 9063.0 9062.5 7181.0 0.02 0.01 1.26 6 48.00 12076.3 12076.7 9346.9 0.02 0.02 1.29 7 60.00 15090.2 15090.4 11590.2 0.02 0.02 1.30 8 80.00 20109.6 20109.9 15157.4 0.01 0.01 1.33 9 100.00 25123.7 25124.2 18633.5 -0.02 -0.02 1.35 10 120.00 30134.4 30133.6 22129.2 -0.05 -0.06 1.36 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1405 to 41213 e- Average count rate measured: 251 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.37 +/- 0.01 e-/ADU RON = 7.01 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.093 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay