*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 12 Jul 2004 15:56:38 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 12 Jul 2004 15:58:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 040712 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.51 stddev = 5.04 FIERA temperature = 290.4 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_040712 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.00 1554.4 1554.6 1267.1 -0.02 -0.02 1.23 2 12.01 3083.0 3083.1 2519.3 -0.02 -0.02 1.22 3 24.02 6142.7 6142.3 4998.6 0.03 0.03 1.23 4 36.04 9204.9 9204.7 7199.3 0.08 0.08 1.28 5 54.05 13793.9 13794.1 10607.8 0.08 0.08 1.30 6 72.07 18382.8 18382.5 14137.1 0.08 0.08 1.30 7 90.09 22969.2 22969.7 17132.7 0.07 0.07 1.34 8 120.12 30600.7 30598.1 22501.5 0.01 0.01 1.36 9 150.15 38205.8 38206.1 26994.6 -0.09 -0.08 1.42 10 180.18 45780.5 45778.3 31508.1 -0.22 -0.22 1.45 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2257 to 66479 e- Average count rate measured: 254 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.45 +/- 0.03 e-/ADU RON = 7.32 +/- 0.13 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.105 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay