*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 21 Jun 2004 15:57:57 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 21 Jun 2004 15:59:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 040621 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.89 stddev = 5.22 FIERA temperature = 290.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_040621 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.01 1557.3 1557.1 1291.8 -0.02 -0.02 1.21 2 12.01 3086.9 3086.7 2462.2 -0.02 -0.03 1.25 3 24.02 6149.1 6149.1 4837.1 0.02 0.03 1.27 4 36.04 9214.8 9214.5 7252.8 0.08 0.08 1.27 5 54.05 13807.8 13808.7 10614.2 0.07 0.08 1.30 6 72.07 18403.0 18402.5 14028.7 0.08 0.08 1.31 7 90.09 22993.1 22992.8 17292.2 0.07 0.07 1.33 8 120.12 30633.3 30632.4 22290.7 0.02 0.02 1.37 9 150.15 38245.1 38244.9 27106.0 -0.09 -0.09 1.41 10 180.18 45829.0 45828.6 31631.7 -0.21 -0.22 1.45 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2255 to 66357 e- Average count rate measured: 255 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.45 +/- 0.02 e-/ADU RON = 7.55 +/- 0.12 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.108 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay