*** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 23 Dec 2003 18:18:00 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 23 Dec 2003 18:19:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 031223 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 265.52 stddev = 5.07 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_031223 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.00 1600.2 1600.5 1286.1 -0.02 -0.01 1.24 2 12.01 3173.3 3173.4 2525.9 -0.02 -0.02 1.26 3 24.02 6322.0 6321.8 4964.7 0.02 0.02 1.27 4 36.04 9475.0 9474.6 7377.3 0.08 0.08 1.28 5 54.05 14198.5 14199.1 10964.5 0.08 0.08 1.29 6 72.07 18923.0 18922.6 14265.3 0.08 0.08 1.33 7 90.09 23645.3 23645.4 17699.5 0.08 0.07 1.34 8 120.12 31498.0 31498.2 22804.2 0.01 0.01 1.38 9 150.15 39327.7 39327.6 27616.1 -0.08 -0.08 1.42 10 180.18 47121.7 47120.9 32247.7 -0.22 -0.23 1.46 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2337 to 68808 e- Average count rate measured: 262 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.46 +/- 0.03 e-/ADU RON = 7.40 +/- 0.13 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.105 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay