*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 07 Sep 2003 15:21:56 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 07 Sep 2003 15:23:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 030907 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 263.81 stddev = 6.46 FIERA temperature = 290.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_030907 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.00 1638.2 1638.0 1327.9 -0.03 -0.02 1.23 2 12.01 3247.6 3247.3 2599.6 -0.02 -0.02 1.25 3 24.02 6468.7 6468.7 5082.9 0.03 0.03 1.27 4 36.04 9694.0 9694.4 7538.2 0.09 0.09 1.29 5 54.05 14526.8 14526.9 11091.2 0.09 0.09 1.31 6 72.07 19359.8 19359.0 14628.0 0.09 0.09 1.32 7 90.09 24188.4 24188.2 18012.5 0.08 0.08 1.34 8 120.12 32220.8 32221.0 23338.9 0.01 0.01 1.38 9 150.15 40225.7 40224.8 28345.2 -0.09 -0.10 1.42 10 180.18 48189.6 48188.4 32713.2 -0.25 -0.25 1.47 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2400 to 70612 e- Average count rate measured: 268 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.47 +/- 0.03 e-/ADU RON = 9.46 +/- 0.17 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.110 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.25 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay