*** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 29 Aug 2003 17:32:33 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 29 Aug 2003 17:33:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 030829 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 272.14 stddev = 5.13 FIERA temperature = 291.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_030829 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.01 1572.2 1572.0 1267.7 0.00 0.01 1.24 2 12.01 3117.8 3117.4 2468.6 -0.06 -0.06 1.26 3 24.02 6214.6 6212.8 4871.6 0.01 -0.02 1.28 4 36.04 9313.3 9312.6 7251.0 0.06 0.05 1.28 5 54.05 13963.9 13962.9 10707.6 0.10 0.09 1.30 6 72.07 18608.7 18607.0 14045.1 0.09 0.08 1.32 7 90.09 23244.6 23253.4 17313.2 0.05 0.09 1.34 8 120.12 30968.4 30984.3 22531.7 -0.00 0.05 1.37 9 150.15 38681.9 38674.7 27251.1 -0.06 -0.08 1.42 10 180.18 46352.1 46341.6 31905.4 -0.19 -0.22 1.45 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2283 to 67292 e- Average count rate measured: 258 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.45 +/- 0.02 e-/ADU RON = 7.44 +/- 0.12 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.096 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay