*** MIDAS logfile (vers. 02FEBpl1.0) *** We, 18 Jun 2003 18:17:43 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** We, 18 Jun 2003 18:18:57 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 030618 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 238.65 stddev = 7.15 FIERA temperature = 291.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_030618 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.01 1632.4 1632.6 1316.0 0.02 0.02 1.24 2 12.01 3234.1 3234.3 2582.5 -0.06 -0.06 1.25 3 24.02 6441.2 6441.8 5013.8 -0.03 -0.03 1.28 4 36.04 9654.0 9654.0 7450.8 0.03 0.03 1.30 5 54.05 14468.7 14468.6 11004.7 0.05 0.05 1.31 6 72.07 19285.2 19286.1 14373.3 0.06 0.07 1.34 7 90.09 24101.7 24101.2 17801.2 0.08 0.07 1.35 8 120.12 32115.7 32114.9 22945.0 0.04 0.04 1.40 9 150.15 40106.1 40105.3 27888.5 -0.04 -0.04 1.44 10 180.18 48059.9 48060.2 32583.9 -0.16 -0.16 1.47 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2410 to 70938 e- Average count rate measured: 267 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.48 +/- 0.02 e-/ADU RON = 10.55 +/- 0.17 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.107 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay