*** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 06 Apr 2003 16:59:02 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 06 Apr 2003 16:59:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 030406 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 246.92 stddev = 6.79 FIERA temperature = 291.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_030406 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.01 1635.2 1635.0 1325.9 0.02 0.02 1.23 2 12.01 3245.3 3245.0 2576.0 -0.07 -0.06 1.26 3 24.02 6469.7 6469.4 5028.6 -0.03 -0.03 1.29 4 36.04 9699.6 9699.1 7452.0 0.04 0.03 1.30 5 54.05 14538.5 14538.5 11103.4 0.04 0.04 1.31 6 72.07 19382.4 19381.7 14699.1 0.07 0.07 1.32 7 90.09 24222.3 24222.8 17634.3 0.07 0.07 1.37 8 120.12 32282.3 32279.2 23162.2 0.05 0.04 1.39 9 150.15 40314.9 40315.1 27711.7 -0.03 -0.03 1.45 10 180.18 48311.4 48311.9 32112.3 -0.16 -0.16 1.50 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2452 to 72434 e- Average count rate measured: 268 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.50 +/- 0.03 e-/ADU RON = 10.18 +/- 0.21 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.084 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay