*** MIDAS logfile (vers. 02FEBpl1.0) *** We, 12 Mar 2003 17:21:33 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** We, 12 Mar 2003 17:22:27 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 030312 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 210.44 stddev = 4.61 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_030312 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.01 1708.7 1708.4 1353.7 0.02 0.02 1.26 2 12.01 3386.4 3386.2 2670.2 -0.05 -0.05 1.27 3 24.02 6745.1 6745.0 5207.4 -0.03 -0.03 1.30 4 36.04 10109.2 10108.6 7722.6 0.02 0.02 1.31 5 54.05 15152.6 15153.0 11391.5 0.04 0.05 1.33 6 72.07 20199.5 20199.3 15085.9 0.07 0.07 1.34 7 90.09 25242.5 25242.4 18458.8 0.07 0.07 1.37 8 120.12 33637.6 33638.2 23661.0 0.04 0.05 1.42 9 150.15 42008.1 42008.3 28681.1 -0.03 -0.03 1.46 10 180.18 50338.6 50338.4 33326.5 -0.16 -0.16 1.51 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2578 to 75957 e- Average count rate measured: 280 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.51 +/- 0.03 e-/ADU RON = 6.96 +/- 0.14 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.102 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay