*** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 27 Jan 2003 22:02:25 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 27 Jan 2003 22:03:19 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 030127 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 241.46 stddev = 2.61 FIERA temperature = 291.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_030127 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.01 1705.9 1706.0 1356.7 0.02 0.04 1.26 2 12.01 3384.7 3384.4 2640.6 -0.07 -0.06 1.28 3 24.02 6746.0 6745.5 5194.4 -0.04 -0.05 1.30 4 36.04 10112.0 10112.4 7737.8 0.01 0.01 1.31 5 54.06 15161.7 15161.3 11337.9 0.04 0.05 1.34 6 72.07 20211.5 20210.0 14900.0 0.07 0.06 1.36 7 90.09 25260.7 25259.8 18337.2 0.08 0.07 1.38 8 120.12 33664.2 33663.7 23810.5 0.05 0.05 1.41 9 150.15 42046.3 42044.3 28876.2 -0.02 -0.02 1.46 10 180.18 50386.1 50383.9 33425.7 -0.14 -0.15 1.51 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2562 to 75681 e- Average count rate measured: 280 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.50 +/- 0.03 e-/ADU RON = 3.92 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.086 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay