*** MIDAS logfile (vers. 02FEBpl1.0) *** Tu, 07 Jan 2003 15:41:23 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Tu, 07 Jan 2003 15:42:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 030107 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 243.27 stddev = 2.42 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA1_030107 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 6.01 1724.0 1723.5 1344.0 0.03 0.03 1.28 2 12.01 3420.2 3420.6 2767.0 -0.08 -0.06 1.24 3 24.02 6819.0 6818.7 5243.2 -0.04 -0.04 1.30 4 36.04 10221.8 10221.8 7780.0 0.02 0.02 1.31 5 54.05 15325.3 15324.5 11523.2 0.05 0.04 1.33 6 72.07 20429.0 20430.2 15015.9 0.06 0.07 1.36 7 90.09 25534.3 25534.1 18645.6 0.08 0.08 1.37 8 120.12 34028.4 34027.4 23925.5 0.05 0.05 1.42 9 150.15 42498.3 42499.4 30250.0 -0.02 -0.02 1.40 10 180.18 50925.0 50923.5 33419.7 -0.15 -0.16 1.52 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2575 to 76067 e- Average count rate measured: 283 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.49 +/- 0.03 e-/ADU RON = 3.61 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.084 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay