*** MIDAS logfile (03SEPpl1.2) *** Tu, 21 Jun 2005 16:05:30 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Tu, 21 Jun 2005 16:06:32 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 050621 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 370.15 stddev = 14.98 FIERA temperature = 290.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_rD2_050621 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 768.9 769.2 803.1 -0.01 0.01 0.96 2 2.00 1527.9 1527.9 1357.2 -0.07 -0.07 1.13 3 4.00 3048.3 3048.4 2476.6 0.01 0.00 1.23 4 6.00 4566.0 4566.6 3641.1 -0.02 -0.02 1.25 5 9.00 6845.6 6845.4 5284.9 -0.01 -0.01 1.30 6 12.00 9124.8 9124.0 6923.2 0.00 -0.01 1.32 7 15.00 11404.1 11404.5 8772.5 0.01 0.00 1.30 8 20.00 15204.4 15205.2 11455.4 0.02 0.02 1.33 9 25.00 19005.4 19006.3 14365.6 0.03 0.03 1.32 10 30.00 22805.1 22806.8 16957.8 0.03 0.04 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1043 to 30925 e- Average count rate measured: 760 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.36 +/- 0.01 e-/ADU RON = 20.31 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.015 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.07 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay