*** MIDAS logfile (05SEPpl1.2) *** Fr, 05 May 2006 15:08:27 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 05 May 2006 15:09:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060505 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.05 stddev = 17.25 FIERA temperature = 289.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_rC2_060505 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1711.3 1712.5 1744.8 -0.07 -0.07 0.98 2 2.12 3321.6 3321.9 2938.9 0.14 0.14 1.13 3 4.25 6513.5 6515.1 5276.5 -0.11 -0.10 1.23 4 6.38 9728.6 9729.3 7663.5 0.02 0.01 1.27 5 9.57 14541.0 14541.3 11217.8 0.04 0.05 1.30 6 12.76 19371.6 19372.9 14740.7 0.15 0.16 1.31 7 15.95 24185.9 24183.2 18109.5 0.15 0.14 1.34 8 21.27 32187.3 32187.9 23814.3 0.07 0.07 1.35 9 26.59 40143.6 40143.8 28939.6 -0.08 -0.08 1.39 10 31.91 48034.0 48031.2 34139.6 -0.32 -0.33 1.41 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2435 to 68314 e- Average count rate measured: 1506 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.42 +/- 0.01 e-/ADU RON = 24.53 +/- 0.21 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.077 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.32 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay