*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 16:10:57 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 16:11:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060114 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 375.14 stddev = 8.27 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_rC2_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 873.4 873.3 728.7 0.03 0.05 1.20 2 2.00 1700.6 1700.2 1353.2 -0.04 -0.05 1.26 3 4.00 3352.2 3352.4 2618.4 -0.14 -0.15 1.28 4 6.00 5008.8 5008.7 3852.9 -0.09 -0.08 1.30 5 9.00 7493.5 7493.3 5772.5 -0.04 -0.05 1.30 6 12.00 9981.0 9980.7 7597.6 0.01 0.00 1.31 7 15.00 12467.1 12467.9 9429.5 0.03 0.03 1.32 8 20.00 16616.2 16614.7 12511.7 0.08 0.07 1.33 9 25.00 20759.5 20761.7 15542.9 0.08 0.09 1.34 10 30.00 24902.2 24902.0 18902.8 0.08 0.08 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1163 to 33172 e- Average count rate measured: 828 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.00 e-/ADU RON = 11.01 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.057 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay