| The EMMI fast 2x2 63C Red CCD Page |
|
Derivatives
Standard CCD Test
Long CCD TestFor printing, the above plot is also available in postscript format . The raw data can also be viewed in a tabular format.
*** MIDAS logfile (05SEPpl1.2) *** Mo, 08 May 2006 23:06:51 Page 1
*** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 08 May 2006 23:07:49 Page 2
==========================================================
SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060508 IN fastcd MODE
==========================================================
BIAS
====
5 biases at start test: average = 356.01 stddev = 11.58
FIERA temperature = 289.9 K
There were 0 hot pixels/columns detected
TRANSFER FUNCTION AND LINEARITY DATA
====================================
Table : Std_fastcd_rC2_060508
Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1
-------- --------- --------- --------- --------- --------- --------- ---------
1 1.06 1712.0 1712.8 1439.0 -0.11 -0.10 1.19
2 2.13 3325.1 3325.3 2637.4 0.15 0.15 1.26
3 4.25 6520.8 6520.9 5066.6 -0.08 -0.09 1.29
4 6.38 9737.4 9737.2 7441.9 0.02 0.02 1.31
5 9.57 14554.4 14554.1 11058.2 0.06 0.06 1.32
6 12.77 19388.4 19388.0 14551.2 0.16 0.16 1.33
7 15.96 24202.8 24203.5 18050.2 0.14 0.15 1.34
8 21.28 32210.8 32211.3 24149.0 0.07 0.07 1.33
9 26.59 40167.4 40170.2 29304.1 -0.09 -0.09 1.37
10 31.91 48065.8 48065.2 34378.3 -0.33 -0.33 1.40
-------- --------- --------- --------- --------- --------- --------- ---------
Dynamic range explored: from 2386 to 66966 e-
Average count rate measured: 1507 ADU/sec.
CONV. FACTOR AND RON
====================
C.Fact = 1.39 +/- 0.01 e-/ADU
RON = 16.14 +/- 0.16 e- RMS
SHUTTER ERROR
=============
Average shutter error derived from fractional count rate = 0.076 sec
LINEARITY ERROR
===============
Average linearity error after shutter correct. = + 0.16/-0.33 %
These errors represent maximum deviation in % from the average
count rate after correction for shutter delay
In addition, the previous standard tests are available. Click in the table below to access these files.
In the above table, the postscript files are:
| Column Number | Description |
|---|---|
| 1 | Average of the rows and cols. of the bias image |
| 2 | Combined bias frame of 5 bias images |
| 3 | Histogram of the bias image |
| 4 | Plots of exposure sequence, unedited transfer and raw linearity |
| 5 | Plots of filtered linearity and differential non-linearity in |
| 6 | Filtered transfer curves for both data sets |
Note: filtered means that saturated images or deviating too much from the other data, if present, have been removed.
Below are the results of the long tests carried out by ODT. Be warned that these are large compressed tar files, and are of specialised interest. To download, press the shift key and click on the hypertext link with your middle mouse button.
| Long Tests |
|---|
|
ntt Team CCD Page |
ntt Team Home Page | Last update: 04/07/06 Comments to nttccd@eso.org |
|---|