*** MIDAS logfile (05SEPpl1.2) *** Th, 19 Jan 2006 22:04:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 19 Jan 2006 22:05:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 060119 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 630.48 stddev = 29.67 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_rB1_060119 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 1017.0 1016.9 1832.0 0.24 0.24 0.56 2 8.00 1998.0 1998.1 2469.0 -0.42 -0.42 0.81 3 16.00 3985.5 3985.5 4210.2 -0.11 -0.10 0.95 4 24.00 5969.7 5969.4 5529.1 -0.06 -0.06 1.08 5 36.00 8938.3 8937.9 7878.7 -0.11 -0.11 1.13 6 48.00 11920.8 11920.1 9874.7 -0.02 -0.02 1.21 7 60.00 14903.4 14903.1 11995.3 0.04 0.04 1.24 8 80.00 19869.4 19869.8 15425.5 0.07 0.07 1.29 9 100.00 24856.6 24857.9 18812.7 0.17 0.18 1.32 10 120.00 29830.5 29830.7 22065.7 0.19 0.20 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1444 to 42369 e- Average count rate measured: 248 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.42 +/- 0.01 e-/ADU RON = 42.14 +/- 0.41 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.095 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.42 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay