*** MIDAS logfile (05SEPpl1.2) *** Sa, 07 Jul 2007 17:49:18 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 07 Jul 2007 17:49:32 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 070707 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 570.58 stddev = 4.15 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_31_070707 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.19 2053.8 2054.5 1454.4 -0.06 -0.06 1.41 2 4.38 3970.9 3969.5 2804.3 0.14 0.09 1.42 3 8.75 7779.3 7781.5 5380.0 0.03 0.05 1.45 4 13.12 11595.3 11595.7 8063.4 0.05 0.06 1.44 5 19.68 17307.4 17311.1 12066.4 -0.02 0.00 1.43 6 26.23 23025.7 23025.4 16056.0 0.00 0.01 1.43 7 32.79 28739.3 28740.2 19948.6 -0.03 -0.02 1.44 8 43.72 38272.2 38268.6 26943.8 -0.02 -0.03 1.42 9 54.65 47798.6 47793.9 32846.2 -0.04 -0.04 1.46 10 65.58 57321.9 57317.4 40154.2 -0.05 -0.05 1.43 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2949 to 82284 e- Average count rate measured: 872 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.44 +/- 0.01 e-/ADU RON = 5.96 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.166 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay