*** MIDAS logfile (05SEPpl1.2) *** Th, 28 Jun 2007 17:01:02 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 28 Jun 2007 17:01:17 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 070628 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 570.57 stddev = 3.82 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_31_070628 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.19 2048.0 2046.4 1457.2 -0.08 -0.04 1.41 2 4.38 3962.9 3957.8 2788.7 0.12 0.06 1.42 3 8.75 7767.5 7769.8 5391.5 0.01 0.07 1.44 4 13.12 11581.0 11576.2 8024.5 0.05 0.03 1.44 5 19.68 17287.5 17289.8 11921.5 -0.02 0.00 1.45 6 26.23 23001.8 22999.1 15797.7 0.01 0.01 1.46 7 32.79 28715.0 28704.6 19592.3 -0.00 -0.03 1.47 8 43.72 38230.2 38228.5 26906.1 -0.03 -0.03 1.42 9 54.65 47749.3 47748.5 33584.7 -0.04 -0.03 1.42 10 65.58 57272.2 57267.6 39633.5 -0.03 -0.04 1.45 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2937 to 82159 e- Average count rate measured: 871 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.43 +/- 0.01 e-/ADU RON = 5.48 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.160 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay