*** MIDAS logfile (05SEPpl1.2) *** Fr, 17 Nov 2006 13:19:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 17 Nov 2006 13:20:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061117 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 572.62 stddev = 3.59 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_31_061117 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.19 2189.0 2187.8 1520.1 -0.02 -0.08 1.44 2 4.38 4227.5 4233.4 2916.2 0.01 0.14 1.45 3 8.75 8296.0 8297.1 5821.4 0.03 0.04 1.43 4 13.12 12365.8 12364.7 8688.1 0.05 0.04 1.42 5 19.68 18466.6 18464.2 12987.8 0.02 -0.00 1.42 6 26.23 24563.7 24556.1 17579.7 0.02 -0.01 1.40 7 32.79 30656.2 30657.9 21076.5 -0.02 -0.02 1.45 8 43.72 40824.8 40823.4 27871.0 -0.02 -0.03 1.46 9 54.65 50990.2 50990.5 35191.4 -0.02 -0.03 1.45 10 65.58 61147.9 61144.9 42983.0 -0.04 -0.05 1.42 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3150 to 88015 e- Average count rate measured: 931 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.44 +/- 0.01 e-/ADU RON = 5.17 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.163 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay