*** MIDAS logfile (05SEPpl1.2) *** Fr, 20 Oct 2006 16:07:23 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 20 Oct 2006 16:08:17 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 061020 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 573.63 stddev = 3.58 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_31_061020 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.19 2160.1 2160.3 1524.1 -0.07 -0.07 1.42 2 4.38 4179.3 4180.4 2887.5 0.10 0.11 1.45 3 8.75 8198.3 8200.0 5714.6 0.06 0.07 1.43 4 13.12 12210.8 12216.4 8392.1 -0.00 0.03 1.46 5 19.68 18244.9 18242.6 12801.9 0.01 -0.01 1.43 6 26.23 24263.7 24266.4 16807.0 -0.01 -0.01 1.44 7 32.79 30291.7 30295.1 21206.5 -0.02 -0.02 1.43 8 43.72 40337.5 40338.4 27989.5 -0.02 -0.03 1.44 9 54.65 50392.8 50389.9 34837.2 -0.01 -0.02 1.45 10 65.58 60426.7 60426.7 41293.6 -0.03 -0.04 1.46 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3145 to 87976 e- Average count rate measured: 920 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.46 +/- 0.01 e-/ADU RON = 5.21 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.161 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.07 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay