*** MIDAS logfile (05SEPpl1.2) *** Fr, 08 Sep 2006 14:10:57 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 08 Sep 2006 14:11:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 060908 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 573.15 stddev = 3.54 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_31_060908 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.19 2229.8 2223.7 1583.3 -0.00 -0.01 1.41 2 4.38 4299.6 4294.2 3013.2 0.01 0.01 1.43 3 8.75 8426.6 8429.7 5861.5 -0.02 0.07 1.44 4 13.12 12560.4 12558.5 8883.7 0.03 0.03 1.41 5 19.68 18757.2 18759.4 13162.2 0.01 0.03 1.43 6 26.23 24937.8 24941.0 17393.8 -0.02 -0.02 1.43 7 32.79 31142.4 31140.3 21606.2 0.00 -0.02 1.44 8 43.72 41467.0 41463.1 28508.3 -0.00 -0.03 1.45 9 54.65 51793.2 51793.4 35745.3 -0.00 -0.02 1.45 10 65.58 62116.4 62110.8 42908.5 -0.01 -0.04 1.45 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3235 to 90241 e- Average count rate measured: 945 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.45 +/- 0.00 e-/ADU RON = 5.14 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.166 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay