*** MIDAS logfile (vers. 02FEBpl1.0) *** We, 29 Jan 2003 13:54:10 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** We, 29 Jan 2003 13:54:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 030129 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 573.54 stddev = 3.50 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_31_030129 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.19 1695.8 1700.4 1206.5 -0.07 0.03 1.41 2 4.38 3290.3 3284.7 2267.8 0.11 -0.13 1.45 3 8.75 6464.8 6464.4 4417.3 0.10 0.07 1.46 4 13.12 9635.3 9634.2 6635.6 0.05 0.04 1.45 5 19.68 14392.5 14390.6 10030.7 0.00 0.01 1.43 6 26.23 19141.1 19140.2 13344.0 -0.03 -0.01 1.43 7 32.79 23899.2 23905.3 16726.9 -0.04 0.01 1.43 8 43.72 31829.3 31827.2 21692.3 -0.04 -0.01 1.47 9 54.65 39759.3 39762.2 27026.0 -0.04 0.00 1.47 10 65.58 47687.3 47686.8 32739.5 -0.04 -0.01 1.46 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2486 to 69813 e- Average count rate measured: 726 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.46 +/- 0.01 e-/ADU RON = 5.13 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.150 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay