*** MIDAS logfile (05SEPpl1.2) *** Fr, 11 Jan 2008 13:27:38 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 11 Jan 2008 13:27:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 080111 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 561.93 stddev = 2.40 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_080111 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1886.1 1889.3 674.5 -0.04 0.02 2.80 2 8.63 3713.4 3710.0 1321.8 0.10 -0.05 2.81 3 17.25 7346.4 7349.1 2569.3 -0.02 0.01 2.86 4 25.87 10982.4 10984.7 3841.9 -0.03 -0.01 2.86 5 38.80 16439.7 16440.4 5859.7 -0.01 -0.01 2.81 6 51.73 21894.7 21892.0 7686.1 -0.02 -0.02 2.85 7 64.66 27348.1 27350.5 9484.9 -0.03 -0.01 2.88 8 86.21 36448.3 36446.7 12641.0 -0.00 0.01 2.88 9 107.76 45552.8 45547.2 16086.6 0.02 0.02 2.83 10 129.31 54651.7 54650.2 19377.9 0.03 0.04 2.82 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5352 to 154952 e- Average count rate measured: 422 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.84 +/- 0.02 e-/ADU RON = 6.81 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.164 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.04 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay