*** MIDAS logfile (05SEPpl1.2) *** We, 02 Jan 2008 00:19:58 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 02 Jan 2008 00:20:13 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 080102 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 561.75 stddev = 2.52 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_080102 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1896.3 1896.3 676.3 0.01 0.01 2.80 2 8.63 3726.3 3726.7 1331.0 -0.03 -0.01 2.80 3 17.25 7381.4 7379.6 2603.9 0.00 -0.02 2.83 4 25.87 11033.4 11034.0 4060.0 -0.02 -0.01 2.72 5 38.80 16514.7 16514.3 5728.6 -0.01 -0.01 2.88 6 51.73 21992.8 21993.1 7766.3 -0.02 -0.02 2.83 7 64.66 27473.8 27472.7 9586.8 -0.02 -0.02 2.87 8 86.21 36613.1 36612.4 12733.9 -0.00 -0.00 2.88 9 107.76 45764.7 45760.8 16032.0 0.04 0.03 2.85 10 129.31 54906.5 54907.0 19160.8 0.04 0.05 2.87 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5449 to 157762 e- Average count rate measured: 424 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.87 +/- 0.01 e-/ADU RON = 7.23 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.163 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.04/-0.02 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay