*** MIDAS logfile (05SEPpl1.2) *** Su, 09 Dec 2007 14:48:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 09 Dec 2007 14:48:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 071209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 561.31 stddev = 2.99 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_31_071209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.31 1904.3 1907.8 676.2 -0.03 -0.02 2.82 2 8.63 3746.5 3747.9 1309.8 0.06 0.03 2.86 3 17.25 7417.1 7416.4 2660.5 0.04 0.01 2.79 4 25.87 11081.9 11083.4 3880.5 -0.02 -0.01 2.86 5 38.80 16586.1 16584.5 5808.3 -0.02 -0.02 2.86 6 51.73 22086.9 22085.9 7748.4 -0.03 -0.03 2.85 7 64.66 27590.9 27593.4 9711.0 -0.03 -0.01 2.84 8 86.21 36766.9 36765.1 13217.3 -0.02 -0.01 2.78 9 107.76 45958.6 45954.4 16061.4 0.02 0.03 2.86 10 129.31 55133.5 55129.1 19040.6 0.02 0.03 2.90 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 5471 to 158249 e- Average count rate measured: 426 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.87 +/- 0.02 e-/ADU RON = 8.58 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.168 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay